Plasma Cleaner

OIL-BASED VACUUM PUMPS

 
Our Plasma Cleaners require a vacuum pump with a minimum pump
speed of 1.4 m3/hr (23 L/min) and an ultimate total pressure of
200 mTorr (0.27 mbar) or less.
 

Rotary vane pumps with hydrocarbon pump oil are suitable for pumping 

nonreactive or inert gases (e.g. Air, N2 or Ar). These pumps are not intended 

for pumping oxygen; see our  Oxygen Service Pumps. For corrosive or 

reactive gases, consult with us directly.

 
BASIC OIL-BASED VACUUM PUMP

PDC-VPE (115V) | PDC-VPE-2 (230V)

 
Basic Model (lower cost)Pumping Speed (50/60 Hz): 5.1 m3/hrUltimate
Total Pressure (gas ballast closed): 7.5 mTorrUltimate Total Pressure
(gas ballast open): 150 mTorrMotor Power (50/60 Hz): 373 WWeight: 27 lbs
Size: 9.75″ H x 14″ W x 5.5″ DPump inlet: includes check valve, hose,
hose clamps, and hose adaptor to connect plasma cleaner to pump
inletPump outlet: includes oil mist filter to filter oil vapor from pump
exhaustPDC-VPE-OIL (1 quart) Replacement pump oil

OXYGEN SERVICE VACUUM PUMPS

ECONOMY DRY OXYGEN PUMP

 
Our Plasma Cleaners require a vacuum pump with a minimum pump
speed of 1.4 m3/hr (23 L/min) and an ultimate total pressure of 200 mTorr
(0.27 mbar) or less.
 
Suitable for pumping concentrated oxygen gas as well as nonreactive or 
inert gases (e.g. Air, N2, Ar).Does not use 
hydrocarbon oil  (the combination
 of concentrated oxygenand hydrocarbon oil is an explosion hazard).
 
Uses no oil or fluidPumping speed (50/60 Hz): 3.0/3.6 m3/hr
Ultimate total pressure: 250 mTorr (pump manufacturer
specification; in our lab tests, ultimate pressures of
<100 mTorr were attained)Motor power (50/60 Hz):
120 WWeight: 21 lbsDimensions: 7″ H x 14″ W x 5.5″ D
Pump inlet: includes hose, hose clamps, hose adaptor,
centering ring and swing clamp to connect plasma cleaner
to pump inle PDC-OPE-OLTOptional pump outlet connection
to conduct pump exhaust to a vent hoodIncludes 9 ft. length
exhaust hose (other lengths may be specified)

 

OXYGEN SERVICE VACUUM PUMPS

DRY OXYGEN PUMP

 
Our Plasma Cleaners require a vacuum pump with
a minimum pump speed of 1.4 m3/hr (23 L/min) and
an ultimate total pressure of 200 mTorr (0.27 mbar) or less.
 
Suitable for pumping concentrated oxygen gas as well as
nonreactive or inert gases (e.g. Air, N2, Ar).Does not
use hydrocarbon oil (the combination of concentrated
oxygen and hydrocarbon oil is an explosion hazard).
 

Uses no oil or fluidPumping speed (50/60 Hz): 6.2 m3/hr

Ultimate total pressure (gas ballast closed; 50/60 Hz): 

15 mTorrUltimate total pressure (gas ballast open; 50/60 Hz): 

38 mTorrMotor power (50/60 Hz): 260 WWeight: 58 lbs

Dimensions: 12″ H x 17″ W x 11″ DPump inlet: includes hose, 

hose clamps, hose adaptor, centering ring and swing clamp to 

connect plasma cleaner to pump inlet PDC-OP-OLTOptional 

pump outlet connection to conduct pump exhaust to a vent 

hoodIncludes 9 ft. length exhaust hose (other lengths may be specified)

 